EasyManua.ls Logo

Abbott i-STAT Alinity - Adjustments

Abbott i-STAT Alinity
296 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adjustments
The sengs for result displays of base excess and eGFR (Esmated Glomerular Filtraon Rate) are
determined in this tab.
BE Equaon Base excess of the extracellular uid or standard base excess is the
concentraon of tratable base minus the concentraon of tratable
acid when trang the average intracellular uid (plasma plus intersal
uid) to an arterial plasma pH of 7.40 at PCO
2
of 40 mmHg at 37°C.
Excess concentraon of base in the average ECF remains virtually
constant during acute changes in the PCO
2
and reects only
nonrespiratory component of pH disturbances. Select either opon:
Extra Cellular Fluid
Base Excess of Extracellular Fluid (BEecf)
BEecf = HCO
3
– 24.8 + 16.2 (pH – 7.4)
This is the default.
Blood
Base Ex
cess of Blood (BEb)
BEb = (1 - 0.014*Hb) * [HCO
3
- 24.8 + (1.43 * Hb
+ 7.7) * (pH - 7.4)]
eGFR E
quaon MDRD (default)
CKD-EPI 2009
eGFR Variants Select one of these opons:
Display both eGFR and Black/African American eGFR
Display only eGFR
Display only Black/African American eGFR
Note: Black/
African American eGFR displays as eGFR-a on the
instrument screen.
When nished, choose fr
om the opons shown at the boom of the screen:
Previous to go back to the previous screen
Discard to undo changes
Finish Later to create a dra
Next to open the next tab
188
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022

Table of Contents