1.7 - Perform Electronic Simulator Tesng
Although the instrument performs internal electronic checks and calibraon during each test cycle, the
Electronic Simulator test provides an independent check on the ability of the instrument to take accurate
and sensive measurements of voltage, current and resistance from the cartridge. The instrument will
pass or fail this electronic test depending on whether or not it measures these signals within limits
specied in the instrument soware.
Electronic Simulator
The Electronic Simulator is a quality control device used to evaluate the iSTAT Alinity instrument’s ability
to read the electronic signal from a cartridge. The test cycle for the Electronic Simulator is approximately
60 seconds.
Relave humidity
The Electronic Simulator test will fail if high humidity interferes with the measurements. It is therefore
unnecessary to record humidity where the instruments are in use.
Storing the Electronic Simulator
Store the Electronic Simulator with the cap on in the box in which it was shipped.
59
i-STAT Alinity — System Operaons Manual Art: 745578-01 Rev. C Rev. Date: 04-Mar-2021

Table of Contents

Other manuals for Abbott i-STAT Alinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT Alinity and is the answer not in the manual?

Abbott i-STAT Alinity Specifications

General IconGeneral
BrandAbbott
Modeli-STAT Alinity
CategoryMedical Equipment
LanguageEnglish

Related product manuals