Printout Text System Administrator Acon
Printer Failure. Use another
printer. Report failed
printer to System
Administrator. 004
Reset the failed printer:
1. Turn o the printer and unplug the power cord.
2. Open the baery compartment and unplug the baery.
3. Leave the baery unplugged for at least 10 seconds.
Then reconnect the baery and close the baery
compartment.
4. Reconnect the power cord. Then, press the power
buon to turn the printer on. Aempt to print a record
from the instrument. If a printer failure occurs, contact
Technical Support.
Cleaning and Disinfecng
Complete instrucons for cleaning and disinfecng are found in the Quick Reference Guide. The graphics
below are a visual representaon of the wiping procedure for disinfecon.
69