The features in this secon determine the acons for Operators when working in the paent test
pathway or training mode pathway.
Operator Acons
Accept or Reject Results Determines whether the operator has the ability to accept or reject
results aer they are displayed on the instrument.
Enable Accept or Reject
Results
Select the check box to allow operators
to accept or reject test results aer
results display on the instrument. If
results are rejected no paent
treatment should be performed from
the rejected results. Unselect the check
box and operators will not have the
opon of rejecng or accepng results
aer they display on the instrument. If
results are transmied to the data
management system, only those that
are accepted will go through to the LIS/
EMR. Rejected results will not be sent to
the LIS/EMR.
Rejecon Comment Select either opon:
Oponal The operator is prompted to enter a
comment for rejecng test results.
Mandatory The operator is prompted to enter a
mandatory comment for rejecng test
results.
Rejecon Comment Style Controls display of comments on the iSTAT Alinity. Select either:
Comment List Only comments chosen from the
selecon list will be displayed and
available for aaching to a rejected
result.
Comment List with text
box
Comments chosen from the comment
list along with a text box that will accept
20 characters will be displayed and
available for aaching to a rejected
result. A maximum of 6 comments can
be displayed with a text box.
154
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022