EasyManua.ls Logo

Abbott i-STAT Alinity - Calibration Verification (Cal Ver)

Abbott i-STAT Alinity
296 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1.11 - Calibraon Vericaon (Cal Ver)
Calibraon Vericaon is a procedure performed to conrm that the calibraon of an instrument or test
system has remained stable throughout the reportable range. This procedure is also called a linearity
check.
Note: Only iSTAT Calibraon Vericaon and iSTAT TriControls Calibraon Vericaon
soluons can be used with a non-customized instrument. See the AlinIQ CWi secon of
this manual for instrucon on creang and assigning customizaon proles to
instruments.
The following four items are reasons to perform calibraon vericaon:
1. Validate the reportable range of a test before the test system is put into use.
The accuracy of results over the enre reportable range could be assessed by tesng the same
paent samples on the new system and on a system with known accuracy and comparing results
using an acceptable dierence criteria.
The target values have been determined over many lots of cartridges and results on these soluons
when compared to the target values indicate the performance of a parcular lot of cartridges.
2. Verify that a change in reagent lot numbers does not aect either the reportable range or control
values.
Lot-to-lot variaon over the enre reportable range for any reagent system could be assessed by
tesng calibraon vericaon soluons on old and new lots in parallel. Quality controls samples with
concentraons at decision points should always be used to assess new lots of reagent before results
are reported.
3. Verify that results have not been aected by maintenance or repair procedures.
The user cannot perform any maintenance procedures on the iSTAT Alinity System. The soware in
the instrument is updated periodically. Calibraon vericaon soluons could be tested to verify that
the system performs as before the upgrade.
Repaired and newly purchased instruments are received with factory calibraon. Tesng calibraon
vericaon samples or comparing paent sample results on a new or repaired instrument with an
older instrument will assess cartridge performance. The Electronic Simulator, rather than calibraon
vericaon or control soluons, provides beer assurance that the instrument is funconing
correctly. Any variaons in instrument performance will not be stascally discernable above the
performance of the cartridges. When mulple instruments are to be used at a facility, Abbo Point of
Care Inc. recommends including at least two instruments in any performance vericaon studies so
that stascs reect the “system.
4. Troubleshoot when control values are out-of-range.
Should quality control sample results fall outside of the acceptable ranges, the use of calibraon
vericaon samples with very low or very high concentraons could be helpful in characterizing a
reagent problem. The characteriscs of the sensors and results of control soluons are sucient for
Technical Support specialists to help users resolve out-of-range control problems.
79
i-STAT Alinity — System Operaons Manual Art: 745537-01 Rev. F Rev. Date: 02-Nov-2022

Table of Contents

Other manuals for Abbott i-STAT Alinity

Related product manuals