*Quality Sengs Category
This secon contains features to customize the quality sengs for the i‑STAT Alinity, such as liquid QC
pass/fail determinaon, Cal Ver pass/fail determinaon, and cartridge lot QC. The Quality Sengs
category can be created or modied with the following steps.
Click Quality Sengs, then select APOCQualitySettings and click View. The screen displays the
default name APOCQualitySettings, and default values for each feature. Scan through the
APOCQualitySettings and determine if the selecons work for the prole being assembled. If they
do then no addional work is needed in this category. The APOCQualitySettings category will be
available in the Manage Proles secon. Use it to assemble the new prole. If the
APOCQualitySettings do not work for the prole being assembled, click Finish to stop viewing.
Click Create and follow the instrucons below.
Note: All r
equired categories have a default seng named with the APOC prex. In
addion, the sengs are populated with default values.
You are prompted to specify a name for the sengs and an oponal descripon:
Quality Sengs
Name Required. Specify a name that is from 4 to 40 characters long and:
• c
onsists only of the following characters:
○ 0 through 9
○ A through Z or alphabet appropriate for language (in upper or
lower case)
○ blank space
○ underscore ( _ )
• does not begin with the characters APOC (in upper or lower
case)
• is a unique name within this healthcare organizaon
Descripon Oponal. Pr
ovide any informaon that helps the user know how or
why the category was named or created.
When nished, choose fr
om the acons shown at the boom of the screen:
• Discard to undo changes
• Next to open the next tab
191