Specicaons
i-STAT Alinity Rechargeable Baery Specicaons
Dimensions: Height x Width x Depth 2.7 in (6.9 cm) x 3.6 in (9.1 cm) x 1.9 in (4.8 cm)
Weight 0.4 lb (180 g)
Electrical rang 3.65 VDC 19.3 Wh (nominal)
Operaonal range: Temperature and humidity 10 to 40ºC (50 to 104ºF)
10 to 90% non-condensing relave humidity, with
maximum saturaon temperature of 34ºC (93.2ºF)
Storage range: Temperature and humidity -10 to 60ºC (14 to 140ºF)
10 to 90% non-condensing, with maximum saturaon
temperature of 50ºC (122ºF)
Altude up to 3,048 meters (10,000 feet)
Precauons and Limitaons
Baeries have been designed to provide a safe, high-capacity power source in a relavely small
lightweight package. However, if misused or abused, these baeries can be dangerous. Follow these
guidelines for the safe handling, use and disposal of the rechargeable baeries.
DO:
refer to the Geng Started Guide for proper charging instrucons.
charge a new rechargeable baery pack for 4 hours before inial use. A fully discharged baery will
be 100% charged and ready for use aer 4 hours.
use only a rechargeable baery pack purchased from Abbo Point of Care.
use only the accessories and consumables specied or supplied for this system by Abbo Point of
Care.
keep a spare, charged baery on hand at all mes.
store charged baery in the original packaging.
check with authories for local, state, and/or naonal requirements for disposal or recycling
requirements for Lithium-Ion baeries.
DO NOT:
dismantle, open or shred the baery.
expose the baery to heat or re. Avoid storage in direct sunlight.
short circuit the baery. Do not store baeries in such a manner that they may short circuit each
other, or allow metal objects to touch the baery contacts.
subject the baeries to mechanical shock.
open, disassemble, modify, crush, burn, shred, or expose to high temperatures. Toxic gases and
liquids are present in baeries.
place the baery in an oxygen-enriched atmosphere.
make any unauthorized repairs or modicaons to this product as this may cause personal injury or
damage to the unit.
48
i-STAT Alinity — System Operaons Manual Art: 745529-01 Rev. G Rev. Date: 04-Mar-2021