Ranges
This secon allows for the seng of ranges. Range opons are reportable, reference and acon. If a
reference and reportable range are set for an analyte, the system denes the area in between these as
abnormal. Seng ranges changes the appearance of the result screen on the instrument. Results that
fall in the reference range have no color or arrows associated with that result (in the result bubble).
Results that fall in the acon range (crical range) will have a red p and a solid arrow in the result
bubble. Arrow direcon indicates if the result is out of range high or low. Results that fall outside of the
reference range, but not yet in the acon range are abnormal. The result bubble will have a yellow p
and hollow arrows indicang whether the result is high or low.
Range
Range opens to a list of analytes and the custom reportable range display. To begin eding:
1. Click on the + icon found in far le hand column.
2. Click on the Edit buon.
3. To customize reference range(s), click on Add Range under the Reference Range secon. This area
allows for ranges to be set based on sample type and/or age and/or sex. It is also possible to enter
only the low and high range independent of sample type, age, or sex.
To make reference ranges sample type dependent, check the box in the Apply column and choose
the sample types applicable by clicking the down arrow in the Sample Type column. The sample
types displayed are either the default types or the types customized in the Sample Types secon
of the Analyte Sengs category.
Reference ranges are age dependent by default. Up to six dierent sets of ranges can be age
dependent. Enter the low and high age limits and units; or, to disable age dependent reference
ranges, click to un-check the box in the Apply column. To use this feature, another customizaon
item must be added to the prole: Paent Sengs > Paent Informaon and Posive Paent
Idencaon > Age and Sex
To make reference ranges sex dependent, check the box in the Apply column and choose the sex
applicable by clicking the down arrow in the Sex column. To use this feature, another
customizaon item must be added to the prole: Paent Sengs > Paent Informaon and
Posive Paent Idencaon > Age and Sex
To customize acon range(s), click on Add Range under the Acon Range secon. The same structure is
in place for changing acon range as it is for reference range.
182
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022

Table of Contents

Other manuals for Abbott i-STAT Alinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT Alinity and is the answer not in the manual?

Abbott i-STAT Alinity Specifications

General IconGeneral
BrandAbbott
Modeli-STAT Alinity
CategoryMedical Equipment
LanguageEnglish

Related product manuals