EasyManua.ls Logo

Abbott i-STAT Alinity - Page 223

Abbott i-STAT Alinity
296 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Below is an example of reference, abnormal and acon ranges. In this example, the reference range was
set as 4.0-7.5. The low acon range was set at 2.0 and the high acon range was set at 15.0. The
instrument calculates the abnormal range. When ranges are set and the prole is loaded onto the
instrument, the result screen will reect these changes. An example of a result screen displaying range
customizaon can be found in Interpreng Results with customizaon features applied.
Note: Ne
ver enter reference, acon or custom reportable ranges outside of the
manufacturers reportable range.
Custom Reportable Ranges
Restrict the display range to custom values. For example, healthcare organizaon managers may wish to
limit the display ranges to values that have been veried using calibraon vericaon materials. Set the
low and high values. Narrowing the reportable range of certain tests may aect the presentaon of
other dependent test results. In the table below, if any of the results in the rst column are outside the
reportable range, the dependent test results listed in the second column will be suppressed (displayed as
<>).
Note: Ne
ver enter a null (blank) value for the Custom Reportable range for Low or High.
The values can be le as -99999.9 for Low or 99999.9 for High.
Note: Rang
e customizaon may not apply to all cartridge types.
183
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022

Table of Contents

Other manuals for Abbott i-STAT Alinity

Related product manuals