Table 3–1: Reportable Range Limitations
Test (Outside the Reportable Range) Dependent Suppressed Tests (displayed as <>)
Na K, Cl, BUN, Anion Gap, Hgb, Hct
Hct Cl, Bun, Anion Gap, Hgb
PCO
2
TCO
2
, Anion Gap, Base Excess, HCO
3
, sO
2
pH TCO
2
, Anion Gap, Base Ex
cess, HCO
3
, sO
2
HCO
3
TCO
2
, Anion Gap, Base Ex
cess, sO
2
Cl Anion Gap
K Anion Gap
TCO
2
Anion Gap
PO
2
sO
2
Click Upda
te to save changes to analyte or click Cancel to discard. Repeat this process for all analytes
needed.
When nished selecng all ranges for all desired analytes, choose from the opons shown at the boom
of the screen:
• Previous to go back to the previous screen
• Discard to undo changes
• Finish Later to create a dra
• Next to open the next tab
184