Issue Soluon
i‑STAT Alinity only displays the last 3 numbers or
leers of the Operator ID or No Operator ID. The
enre Operator ID should display.
Check the Operator ID Presentaon feature in the
Operator ID Entry feature on the prole assigned
to the instrument. Follow CWi instrucons in this
manual to change the display.
i‑STAT Alinity only displays the operator’s rst
name and last inial. The enre name should
display.
Check the Operator Name Presentaon in the
Operator ID Entry feature on the prole assigned
to the instrument. If the check box is marked for
Parally Display Operator Name, uncheck it to
display enre name. Follow CWi instrucons in
this manual.
i‑STAT Alinity will not upload operator list Check and conrm that the Search for Operator ID
on Operator List feature in the Operator Sengs
Category is checked on the prole assigned to the
instrument. If not checked, follow CWi instrucons
in this manual to change the prole and enable
the instrument to load the operator list from the
data management system. Check the Connecvity
Sengs Category to conrm the IP address is
programmed for operator list.
i‑STAT Alinity will not upload cartridge lot list Check to conrm Search Cartridge List for
Cartridge Lot Number is enabled in the Quality
Sengs Category on the prole assigned to the
instrument. If not enabled, follow CWi instrucons
in this manual to enable Search Cartridge List for
Cartridge Lot Number to load the cartridge list
from the data management system. Check the
Connecvity Sengs Category to conrm the IP
address is programmed for cartridge list.
Paent armband barcode will not scan Check Barcode sengs in the Paent Sengs ID
Entry feature on the prole assigned to the
instrument. Follow CWi instrucons in this manual
to change the barcode type.
Paent list will not upload onto the i‑STAT Alinity Check and conrm that the Search Paent List for
Paent ID feature is enabled in the Paent
Sengs Category on the prole assigned to the
instrument. If it is not, follow the CWi instrucons
in this manual to change the prole and enable
the instrument to load the Paent List from the
data management system. Check the Connecvity
Sengs Category to conrm the IP address is
programmed for Paent List.