QC Cartridge Click the arrow to display a drop down list of cartridges. From the list, select
the c
artridge for this prole. This is the parent cartridge.
Dependent Cartridges When more than one type of cartridge can test for the same analyte,
one c
artridge can be the dependent of the other. When tesng is
performed, the result for an analyte common to both cartridges then
applies to both the cartridge that was tested as well as its dependent.
For example, a CHEM8+ cartridge tests for sodium. An E3+ cartridge
also tests for sodium. If CHEM8+ is selected from the drop down list
tled "QC Cartridge", then an E3+ cartridge can be selected as its
dependent. The CHEM8+ is the parent and the E3+ is the dependent.
When QC tesng is done for the CHEM8+ cartridge and the result is a
'pass' for sodium, then the E3+ cartridge is also considered as having
passed the test for sodium.
Opon to select one or more cartridges as dependents. Choose a
dependent by selecng the check box next to the cartridge name.
Note: All analy
tes used within the facility should be included
in liquid QC and Cal Ver tesng schedules.
Note: A QC c
artridge that is selected as a parent cartridge in
one Cartridge QC Prole cannot be a dependent cartridge in
another Cartridge QC Prole within the same schedule. It can
be a dependent cartridge in a Cartridge QC Prole belonging
to a dierent schedule.
QC Fluids Select up to 6 di
erent uids in 1 QC schedule. Selecons must not
be duplicated. Aer uids have been selected, click Update QC
Prole.
202
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022