The Base St
aon and instrument light should be blue and the following baery symbol
in the
upper righ
t-hand corner on the instrument should be displayed.
2. The instrument is running soware version OSi05 or later.
To verify:
From the Home screen, navigate to: More Opons > Instrument Status
Select page that shows the Firmware and verify the name of the rmware is OSi05 or later (i.e.,
OSi06, OSi07, OSi08, etc.).
Procedure:
P
erform the following steps to download and install a soware update using the USB memory device:
CA
UTION: During the installaon, do not remove the instrument or the USB memory device from the
Base Staon. When the installaon is complete, the instrument will display a message indicang that
new soware has been installed. The display of this message is the indicaon that it is safe to remove
the instrument and USB memory device from the Base Staon.
1. From the Home screen, navigate to: More Opons > Instrument Opons > Soware Installaon >
Install From USB
2. Follow the on-screen instrucons.
3. When presented with the name of the soware package to be installed, record the name before
proceeding (this informaon may be used at the end of the procedure to verify that the soware
update has been completed).
Press Next to connue with the installaon of the soware package, OR
Press Cancel to terminate the installaon
4. Connue to follow the on-screen instrucons. During the installaon process, the instrument screen
will go blank and remain blank for several minutes. This is normal.
Do not remove the instrument from the Base Staon.
Do not remove the memory device from the Base Staon.
5. When the installaon is complete, the instrument screen will illuminate and display a message that
new soware has been installed. The display of this message is the indicaon that the installaon is
complete and it is safe to remove the instrument and USB memory device from the Base Staon.
98
i-STAT Alinity — System Operaons Manual Art: 754496-01 Rev. E Rev. Date: 02-Nov-2022

Table of Contents

Other manuals for Abbott i-STAT Alinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT Alinity and is the answer not in the manual?

Abbott i-STAT Alinity Specifications

General IconGeneral
BrandAbbott
Modeli-STAT Alinity
CategoryMedical Equipment
LanguageEnglish

Related product manuals