A. From the Home screen, navigate to: More Opons > Instrument Opons > Network Sengs >
Network Services
B. Enter your Operator ID and sequence through the workow to view the press Next to view the
Firmware Delivery opon. Verify Enabled is selected.
5. The baery charge level is no less than 51%.
To verify:
A. From the Home Screen, navigate to: More Opons > Instrument Status
B. Locate Baery and verify level is 51% or higher.
Procedure:
Perform the following steps to download and install soware update:
CA
UTION: During the installaon, do not remove the instrument from the Base Staon. When the
installaon is complete, the instrument will display a message indicang that new soware has been
installed. When installaon is complete, the instrument may be removed from the Base Staon.
1. From the Home screen, navigate to: More Opons > Instrument Opons > Soware Installaon >
Check for Update
2. Aer the soware download is completed, installaon of soware can begin. Please skip step 3.
3. Soware download and installaon can be performed at dierent mes if needed. The i‑STAT Alinity
will display the following symbol
on the Home screen in the lower right hand corner if
sow
are is downloaded but not installed. Touch this icon to display the Soware Installaon menu.
Touch the Install Pending buon to install the soware.
4. Follow the on-screen instrucons to connue with soware installaon.
5. When presented with the name of the soware package to be installed, record the name before
proceeding (this informaon may be used at the end of the procedure to verify that the soware
update has been completed).
• Touch Next to connue with the installaon of the soware package, OR
• Touch Cancel to terminate the installaon
6. Connue to follow the on-screen instrucons. During the installaon process, the instrument screen
will go blank and remain blank for several minutes. This is normal.
• Do not remove the instrument from the Base Staon
7. When the installaon is complete, the instrument will display a message indicang that new soware
has been installed. The display of this message is the indicaon that it is safe to resume normal
operaon of the instrument.
102
i-STAT Alinity — System Operaons Manual Art: 754496-01 Rev. E Rev. Date: 02-Nov-2022