Symptom Acon to take
Check for Update buon is not enabled Verify the following as described in the
prerequisites, then repeat the soware update
procedure:
The instrument is running soware version
OSi05 or later
Network Services for soware update enabled
on instrument
The baery charge level is no less than 51%
Instrument indicates a c
onnecon could not be
made to the Server service, OR
Instrument powers o before compleng the
soware download, OR
Instrument does not power on within 10 minutes
aer screen going blank
Verify the following as described in the
prerequisites, then repeat the soware update
procedure:
i-STAT Alinity instrument(s) with appropriate
Network Conguraon le installed
Base Staon with power cable connected to AC
mains power
Ethernet Cable connected to healthcare
organizaon network
The healthcare organizaon network is
allowing the instrument to connect to the
internet in order to have access to the Abbo
Managed Server or SDi.
Repeated aemp
ts to update soware via wired
network connecon fail
Follow the procedure to perform soware update
using USB memory device
104
i-STAT Alinity — System Operaons Manual Art: 754496-01 Rev. E Rev. Date: 02-Nov-2022

Table of Contents

Other manuals for Abbott i-STAT Alinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT Alinity and is the answer not in the manual?

Abbott i-STAT Alinity Specifications

General IconGeneral
BrandAbbott
Modeli-STAT Alinity
CategoryMedical Equipment
LanguageEnglish

Related product manuals