Symptom Acon to take
Check for Update buon is not enabled Verify the following as described in the
prerequisites, then repeat the soware update
procedure:
• The instrument is running soware version
OSi05 or later
• Network Services for soware update enabled
on instrument
• The baery charge level is no less than 51%
Instrument indicates a c
onnecon could not be
made to the Server service, OR
Instrument powers o before compleng the
soware download, OR
Instrument does not power on within 10 minutes
aer screen going blank
Verify the following as described in the
prerequisites, then repeat the soware update
procedure:
• i-STAT Alinity instrument(s) with appropriate
Network Conguraon le installed
• Base Staon with power cable connected to AC
mains power
• Ethernet Cable connected to healthcare
organizaon network
• The healthcare organizaon network is
allowing the instrument to connect to the
internet in order to have access to the Abbo
Managed Server or SDi.
Repeated aemp
ts to update soware via wired
network connecon fail
Follow the procedure to perform soware update
using USB memory device