DO NOT:
• aempt to remove a cartridge during the tesng cycle. The force that would be necessary to do so
could damage the instrument. The message Cartridge Locked in Instrument. Do Not
Attempt to Remove the Cartridge. remains on the screen unl the instrument unlocks
the cartridge.
• use the instrument in environmental condions that exceed the operang temperature and humidity
specicaons.
• make any unauthorized repairs or modicaons to this product as this may cause personal injury or
damage to the unit.
NOTE:
• Protecon provided by this equipment may be impaired if used in a manner not specied by Abbo
Point of Care.
• Operators should use standard precauons whenever handling the instrument, cartridge, and
peripherals to protect themselves from blood-borne pathogens. Standard precauons, such as the
wearing of gloves, are designed to protect personnel from blood-borne pathogens and pathogens
from other body substances. These precauons are based on the assumpon that blood, body uids,
and ssue can contain infecous agents and should therefore be treated as biohazardous materials.
For more informaon, refer to the CDC/NIH manual Biosafety in Microbiological and Biomedical
Laboratories, 4th edion, 1999, or to the WHO Laboratory Biosafety Manual, 2nd edion, 2003.
• The instrument can be rendered inoperave by damage caused by mishandling -- for example,
dropping the instrument.
• Clinical sengs that demand fail-safe tesng should have a backup instrument or test source
available.
• The instrument and its peripherals are not listed by any authority with respect to suitability for use in
oxygen-enriched atmospheres.
• Synchronizing the instrument date/me to a data manager with an incorrect date/me could render
the instrument unusable. To disable automac date/me synchronizaon, run the Set Clock ow and
uncheck the box Synchronize Clock with Data Manager. Then press Set Date/Time Manually and set
the correct date/me.