Chapter 10 159
Command Reference
:CORR:DATA
10. Command Reference
• When specifying STANdard3 as parameter 1:
In any case, if the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if the upper
limit of the range is exceeded) is set.
MAX or MIN can be used to specify parameter 2 and parameter 3.
Query response {numeric 1},{numeric 2}<newline><^END>
Related commands :CORR:OPEN on page 165
:CORR:SHOR on page 165
:CORR:LOAD on page 160
:CORR:MULT on page 161
:CORR:MULT:CHAN on page 161
:CORR:CKIT:STAN1:FORM on page 154
:CORR:CKIT:STAN2:FORM on page 154
:CORR:CKIT:STAN3:FORM on page 156
:FORM on page 182
Equivalent key
sequence
[Open]( [Blue] - [4] ) - MeasVal
[Short]( [Blue] - [5] ) - MeasVal
[Load]( [Blue] - [6] ) - CorVal - MeasVal
You can only check the setup values. You cannot set them.
Parameter 2: <numeric 1> Parameter 3: <numeric 2>
Description
Value of the primary parameter
*1
.
Value of the secondary
parameter
*1
.
Range -999.999 to 999.999 -99.9999E9 to 99.9999E9
Initial value 100E-9 0
Unit F (farad) Depends on the type of parameter.
*1.Use the :CORR:CKIT:STAN3:FORM command to specify the type of parameter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals