3-64 PN: 10580-00321 Rev. M S331L UG
3-16 Sweep Menu Cable and Antenna Measurements
3-16 Sweep Menu
Key Sequence: Sweep (3)
Data Points: Sets the number of data points: 130,
259, 517, 1033 or 2065.
Run/Hold: Toggles between Run and Hold. When in
Hold mode, pressing this key starts the sweeping and
provides a trigger. When in the Run mode, pressing
this key pauses the sweep.
Sweep Type: This toggles the sweep between single
sweep and continuous sweep. In single sweep mode,
each sweep must be activated by the Run/Hold key.
RF Immunity, High / Low: The instrument defaults to
RF Immunity High and is the suggested setting. See
“RF Immunity” on page 3-30 for details.
RF Pwr In Hold: Sets the RF Output power to be
turned off when Run/Hold is toggled to Hold. See
“RF Pwr In Hold” on page 3-30 for details.