2-4 PN: 10580-00321 Rev. M S331L UG
2-3 Test Panel Connector Overview Instrument Overview
2-3 Test Panel Connector Overview
The test panel for the Site Master S331L is shown in Figure 2-2.
14 Main Menu Keys
15 Warning and Status Area
16 Short Cut Tool Bar (not available in Classic mode)
17 Measurement Settings Summary (touchscreen menu shortcuts)
1 RF Out/Reflect In port (Type N, Female) – 50 ohm impedance;
depending on hardware revision, maximum input is +42 dBm or
+23 dBm at ±50 VDC. Check connector label. Torque to 12 in·lbs.
This connector port for Cable and Antenna measurements. Before
making measurements, connect a phase-stable test port cable to the
RF Out/Reflect In port connector and perform a calibration using the
internal InstaCal (#2) or an external OSL. Refer to “Calibration
Symbols” on page 2-26 and Chapter 7.
2 InstaCal/Power Meter port (Type N, Male) – 50 ohm impedance;
maximum input is +27 dBm, ±45 VDC. Torque to 12 in·lbs.
This port supports dual functions. It can be used for performing user
calibrations in both Cable & Antenna Analyzer modes, or it can be
used as a power meter in Power Meter mode. It is not capable of
performing both functions simultaneously.
Figure 2-2. S331L Test Panel Connector (1 of 2)
Figure 2-1. Site Master Instrument Overview (2 of 2)