S331L UG PN: 10580-00321 Rev. M A-1
Appendix A — Instrument
Messages
A-1 Introduction
This appendix provides additional details regarding messages displayed
on the S331L SiteMaster. The text of the dialog boxes are shown below
and are listed in alphabetical order.
Example Message:
Message Shown in Instrument: Additional details or suggestions
regarding the message.
1. Command to USB device returns error or invalid data: Check the
connection to the USB device and try again.
2. Device not ready: Device being accessed is not ready or
responding. Try to access again.
3. EEPROM corrupted: EEPROM device being accessed has been
corrputed. Perform a Factory Reset under the “Preset Menu”
on page 9-23 and then cycle the instrument power. If the error
m e s s a g e p e r s i s t s , c a l l y o u r A n r i t s u S e r v i c e C e n t e r .
4. Encountered error in loading cable list. Default file restored:
The cable list file has been corrupted and will not be loaded. The
default file has been used.
5. Encountered error in loading keyboard settings. Default file
restored: The onscreen keyboard file has been corrupted and will
not be loaded. The default file has been used.
6. Encountered error in loading shortcut icon settings. Default file
restored: The menu shortcut file has been corrupted and will not
be loaded. The default file has been used.
7. Encountered error in restoring last setup. Default file restored:
The last setup file has been corrupted and will not be loaded.
The default file has been used.