3-28 PN: 10580-00321 Rev. M S331L UG
3-6 Measurement Setup Cable and Antenna Measurements
3-6 Measurement Setup
This section describes how to set up the Cable and Antenna parameters,
markers, and limit lines.
Frequency
Setting up the Measurement Frequency using Start and Stop
Frequencies
1. Press the Freq/Dist main menu key then Frequency if the menu is
collapsed.
2. Press the Start Freq submenu key and enter the start frequency
using the Up/Down arrow keys, rotary knob or keypad. When
using the keypad, the button labels change to GHz, MHz, kHz,
and Hz. Press the appropriate unit key to complete the entry.
3. Press Stop Freq and enter the stop frequency.
Distance
1. Press the Freq/Dist main menu key and then Distance if the menu
is collapsed.
2. Press the Start Distance submenu key and enter the start distance
using the Up/Down arrow keys, rotary knob or keypad. When
entering a distance using the keypad, press the unit key to
complete the entry.
3. Press Stop Dist and enter the stop distance.
Amplitude
Setting the Amplitude using Top and Bottom Keys
1. Press the Amplitude main menu key.
2. Press the Top submenu key and use the keypad, rotary knob, or
the Up/Down arrow key to edit the top scale value. Press Enter
to set.
3. Press the Bottom key and use the keypad, rotary knob, or the
Up/Down arrow key to edit the bottom scale value. Press Enter
to set.