Chapter 4: Running Samples 147
A statistics view is added to the worksheet.
6 Right-click the statistics view and select Edit Statistics View.
7 Edit the statistics view as follows:
• In the Header tab, select the checkbox to Use 2 columns for display.
• On the Populations tab, deselect the checkboxes for all populations
except FITC, PE, PerCP-Cy5.5, and APC.
• On the Statistics tab, select the Mean checkboxes for the
fluorescence-A parameters.
8 Resize the statistics view so it fits on the page.