Chapter 6: Shutdown and Maintenance 181
Scheduled Maintenance
For optimal cytometer functioning, perform the following procedures according
to the recommended maintenance schedule.
Internal Cleaning
BD FACSDiva software includes five pre-programmed cleaning modes that can
be used alone, or in combination to provide the required level of cleaning.
The following sections describe the
different cleaning modes. See Table 6-1
on page 182 for an overview of each
mode.
Procedure Recommended Frequency
Internal Cleaning on this page See Table 6-1 on page 182
Purging Filters on page 198 Weekly
Changing the disposable waste cap (see
Emptying the Waste on page 114)
Monthly
Changing Fluid Filters on page 199 Every 3–6 months
Changing the Sample Lines on page 200 Every 4–6 months, or when required
Changing the Air Filters on page 203 Every 6–12 months, depending on
cytometer use and the quality of the air
Changing the Bubble Filter on page 203 Every 6 months or when increased
debris appears in the FSC vs SSC dot
plot