166 BD FACSAria User’s Guide
Change the sort precision mode to Purity (two tubes), 4-Way Purity
(four tubes), or Single Cell (plate or slide).
Enter the number of target events by choosing a value from the drop-
down menu or entering a number in the field.
Select Save Conflicts if you are using a 2- or 4-Tube Sort Layout and
want to save conflicts.
Select the sort location field(s) to be sorted into. Select multiple fields
by dragging the mouse. Select a row or column by clicking the row or
column header.
Add the required population(s) to each sort location field.
To display fewer counters in the Sort Layout window, click the View
Counters button and deselect a menu option. The corresponding
counter is hidden. (Only counters with a checkmark next to the name
are displayed.)
Starting and Monitoring the Sort
1 Open the sort collection chamber door and install the collection tubes,
plate, or slide.
2 Install the sample tube onto the loading port and click Load.
3 Adjust the required flow rate, if necessary.
Typically, flow rates less than 6 give the best sorting results.
4 (Optional) Turn on the deflection plates and open the aspirator drawer.
The deflection plates turn off automatically each time a tube is unloaded. If
you do not turn them back on before beginning a sort, a dialog box appears
where you can turn on the plates and open the aspirator drawer by clicking
OK.