Setting Up for Sorting Into a Plate or Slide
The following sections describe how to set up for sorting into a plate or slide. For
general guidelines, see Setting Up for Sorting on page 154.
Installing the Sorting Hardware
1 Install the splash shield below the aspirator drawer.
Close the sort block door and open the sort collection chamber door, if
needed. The sort block door must be closed in order to open the collection
chamber door.
Remove the tube holder, if one is installed, and tilt the splash shield to
insert it into the slotted fittings below the sort aspirator drawer. Push the
splash shield all the way in.
2 Click the Access Stage button to bring the ACDU stage to the front.
Any cytometer surface that comes in contact with biological specimens can
transmit potentially fatal disease. Use universal precautions when handling
sorting hardware. Wear suitable protective clothing and gloves.