EasyManuals Logo

Bruker NMR BSMS System User Manual

Bruker NMR BSMS System
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
L-TRX/L-19F
106 / 234 H172203_1_001
2H-REC IN (J1, SMA):
2H receiver input from 2H preamplifier (HPPR/2, AQS 1H2H, LOCK-OUT)
• Maximum input power with no damage: +0 dBm
REF IN (J3, SMA):
Reference clock input from AV4 REFERENCE board or L-19F REF OUT (J3).
• Nominal input power @ 160 MHz: -0.5 ±0.5 dBm.
• Nominal input power @ 320 MHz: -3.5 ±0.5 dBm.
TP-F0 OUT (J4, SMA):
Transmitter pulse output to HPPR preamplifier (via AV4 PSM-A board).
• 5V TTL output, no damage with 50 ohm load.
(pulse polarity see Service Web [}110])
BLNKTR-2H~ IN (J5, SMA):
Power amplifier blanking pulse from AV4 TRX (via NanoBay backplane or AV4 PSM-A
board).
• 5V TTL input
(high (+5 V) = power amplifier blanked, low (0 V) = power amplifier gated)
SEL-2H/AMP~(J6, SMA):
Power amplifier control signal from AV4 GTU unit (SEL_2H/DEC).
• 5V TTL input
(high (+5 V) = 2H Lock, low (0 V) = 2H amplifier selected for gradient shimming, 2H Lock
muted)
2H IN (J8, SMA):
Power amplifier input RF-signal from AV4 TRX.
• Maximum input power with no damage: +10 dBm.
2H-TR OUT (J9, N):
2H RF-output to HPPR preamplifier (2H TRANSM) or L-19F unit (2H-TR IN) depending on
configuration.
• Transmitter output specification see Technical Data BSMS/2 Lock Transceiver [}99].

Table of Contents