8.6.2 Lock Parameters and Commands
The following dialog provides - alternatively to the TopSpin application - the setup of nucleus
specific parameters and invoking of Lock/Auto Lock functions.
Figure8.7: Basic Field Lock Operations by Service Web
1 All Lock Field relevant parameters can be defined here. The specified Drift value
(Field Units per 24 hours) is applied / compensated while manual Drift
Compensation is active. It is possible to define the behavior for Locking out (Lock
Out Convention). When the default option is selected (Keep Field after Auto
Lock) then the Field value of the locked state remains after lock off only if the
locked state has been reached by the Auto Lock command (e. g. TopSpin
Command Lock and definition of solvent). Alternatively, it is possible to force
resetting the Field to the value it had before it was locked, or the Field value of
the locked state can be kept always after lock off.