VTA
H172203_1_001 183 / 234
VTA BTO
Parameter Min Type Max Unit
Temperature measurement Range -270 400 °C
Thermocouple type T Accuracy +/- 0.5 °C
Number of channels 1
Connector type
a
C
BTO2000 Supply Voltage 24 V
Current 500 mA
Connector type D
Heater Connector type H
Maximum voltage 50 V
Maximum current 7 A
Safety temperature measurement Range -200 +850 °C
Thermocouple type K Resolution 0.1
Measurement update rate 1 s
-1
a
Connector types: See Figure 16.1 [}187]
Table16.3: VTA BTO
VTA CRP
Parameter Min Type Max Unit
Temperature measurement Range -200 +850 °C
PT100 Accuracy, without
sensor
+/- 0.25 °C
Number of channels 1
Connector type
a
M/N
Heater Connector type M/N
Maximum voltage 50 V
Maximum current 2.5 A
Safety temperature measurement Range -200 +850 °C
PT100 Resolution 0.1
Measurement update rate 1 s
-1
a
Connector types: See Figure 16.1 [}187]
Table16.4: VTA CRP