EasyManua.ls Logo

Bruker NMR BSMS System User Manual

Bruker NMR BSMS System
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
BSMS System
for AVANCE NEO
User Manual
Version 001
Innovation with Integrity
NMR

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Bruker NMR BSMS System and is the answer not in the manual?

Bruker NMR BSMS System Specifications

General IconGeneral
CategoryChassis
ManufacturerBruker
ModelBSMS
TypeNMR
CompatibilityNMR spectrometers
DimensionsVaries depending on configuration
WeightVaries by configuration
Power RequirementsVaries by configuration

Summary

Introduction to BSMS System

Intended Use of the BSMS System

Defines the sole purpose of the BSMS system as an electronic subsystem for AVANCE NEO NMR spectrometers.

Bruker's Policy Statement

Outlines Bruker's commitment to product improvement and the right to change specifications.

Installation and Initial Commissioning Procedures

Specifies that installation and commissioning must be done by Bruker Service or authorized personnel.

Limitation of Liability

Details manufacturer's liability exclusions for improper use, unauthorized modifications, etc.

Warranty Terms and Conditions

States that warranty terms are included in the manufacturer's general Terms and Conditions.

Customer Service Information

Provides information on how to contact customer service for technical support.

Product Safety and Electromagnetic Compatibility Standards

Lists safety requirements and standards the device complies with, including IEC and UL.

Safety Guidelines for BSMS Operation

General Safety Precautions

Warns about high voltages and the need to disconnect power before internal work.

System Owner's Responsibility and Obligations

Defines the system owner's duty to ensure safe operation and personnel training.

Personnel Requirements and Qualifications

Specifies requirements for personnel operating and maintaining the device.

Personal Protective Equipment (PPE)

Outlines the use of PPE to protect personnel from potential hazards during operation.

Basic Dangers and Workplace Safety

Details general workplace dangers, electrical hazards, and mechanical risks.

Environmental Protection Guidelines

Provides instructions for handling and disposing of pollutants to prevent environmental damage.

Identification of Signs and Labels

Explains various warning signs and labels used on the device and in the work area.

Transport, Packaging, and Storage

Symbols on Packaging

Explains standard symbols found on packaging for safe handling and transport.

Inspection Upon Delivery

Provides instructions for inspecting received shipments for completeness and transport damage.

Packaging and Handling Materials

Details packaging methods, material usage, and handling of packaging materials.

Storage Conditions for Packages

Specifies recommended conditions for storing packages to maintain integrity.

AV4 BSMS System Overview

Introduction to AV4 BSMS

Provides historical context and evolution of the BSMS system, highlighting key features.

Subunits within the AV4 BSMS Chassis

Details the primary subunits of the AV4 BSMS, including Shim, Lock, and Gradient amplifier.

Software Control of the AV4 BSMS

Explains how the AV4 BSMS subunits are accessed and controlled via a web-based service tool.

Rack Configurations for AV4 BSMS

Describes various standard configurations and compatibility notes for AV4 BSMS chassis.

Board Locations in the AV4 BSMS

Illustrates the physical placement of boards within the AV4 BSMS chassis.

Power Supply Requirements for AV4 BSMS

Outlines the power supply requirements based on system configurations and installed units.

System Architecture and Overview

Presents a functional system architecture diagram showing UI, ELCB, and hardware interactions.

Service Web Access and IP Address

Explains how to access the Service Web and the IP address for the AV4 BSMS.

Chassis (Mainframe) Description

Overview of the BSMS Chassis

Provides a visual overview of the BSMS chassis front view.

Description of BSMS Chassis Components

Details the components housed within the AV4 BSMS chassis, including backplane and power supplies.

Power On/Off Functionality

Explains that the BSMS Chassis has no dedicated power switch; power is managed by the PDU.

Backplane Functions

Lists the functions incorporated by the backplane, such as power distribution and slot ID.

Slot Functions and Unit Assignments

Details the assignment of typical units to specific slots within the BSMS chassis.

Part Numbers for BSMS Chassis

Lists the Bruker part number for the AV4 BSMS CHASSIS.

Technical Data of the BSMS Chassis

Provides technical and mechanical specifications for the BSMS chassis.

Troubleshooting the BSMS Chassis

Explains that the chassis is a FRU and directs users to Bruker Service for replacement.

Fan Tray Maintenance and Troubleshooting

Overview of the AV4 BSMS Fan Tray

Visual overview of the fan tray, identifying its location and purpose.

Description of the Fan Tray

Details the fan tray's location, components, and control functions.

Part Numbers for the Fan Tray

Lists the Bruker part number for the AV4 BSMS FAN TRAY.

Technical Data of the Fan Tray

Provides technical and mechanical specifications for the fan tray.

Troubleshooting the Fan Tray

Explains that the fan tray is a FRU and directs users to Bruker Service for replacement.

Fan Tray Diagnostics via Service Web

Describes how fan status is supervised by ELCB software and viewed on the Service Web.

Power Supply Modules for BSMS

System Overview of BSMS Power Supplies

Visual overview of the different BSMS power supply modules.

Description of the Power Supply System

Explains the basic power supply system configuration using AC/DC and DC/DC modules.

AV4 PSM-48V Power Supply Module

Details the AV4 PSM-48V module, its connectors, status indication, and lack of serviceable parts.

AV4 PSM-B Power Supply Module

Describes the AV4 PSM-B module, its LED status, and the location of its three fuses.

AV4 PCTRL Power Control Module

Details the AV4 PCTRL module, its features, status LEDs, and fuse locations.

Part Numbers for BSMS Power Supply

Lists the Bruker part numbers for the various BSMS power supply modules.

Technical Data Summary of Power Supplies

Provides a summary of output specifications for PSM-48V, PSM-B, and PCTRL modules.

Troubleshooting Power Supply Modules

Addresses troubleshooting for PSMs, including fuse replacement and diagnostics.

ELCB (Extended Lock Control Board) Functions

Introduction to ELCB

Describes the ELCB as a combined board for Digital Lock and BSMS control, with improved performance.

ELCB Lock Parameters and Technical Data

Lists lock parameters, their ranges, factory defaults, units, and notes.

Configuration and Wiring of ELCB

Explains ELCB slot placement and necessary connections for operation with TopSpin.

System Architecture and Overview of ELCB

Presents a block diagram of the ELCB's functional system architecture.

ELCB Protection Mechanisms

Details the protection features for the power amplifier providing lock current.

Lock Software Architecture

Describes the abstract control domain and real-time domain of the lock software.

Lock Control State Machine

Illustrates the state machine that handles lock requests and transitions between lock states.

High Gradient Rates for Auto Shim and Drift Compensation

Explains how ELCB handles gradient rates for auto shim and drift compensation.

Measurements Provided for ELCB Diagnostics

Lists diagnostic measurements performed by the ELCB upon startup and RF board tests.

ELCB Calibration Procedures

Describes user-selectable calibration options for lock field or shift.

Front Panel LEDs During ELCB Start Up

Illustrates the meaning of LEDs on the ELCB front panel during system startup.

Bus Interface and Front Connectors

Details the bus interface and front panel connectors for the ELCB.

Service Software and Lock Service Web

Explains access to service functions via a web interface for setup, calibration, and diagnostics.

ELCB Lock Parameters and Commands

Provides a dialog for setting nucleus-specific parameters and invoking lock functions.

ELCB Lock Configuration Settings

Details configuration settings for NMR lock, including frequency, drift, and phase.

ELCB Part Numbers

Lists the Bruker part number for the ELCB.

ELCB Diagnostic and Troubleshooting

Guides on diagnosing and troubleshooting issues related to the ELCB and lock system.

SCB20 Shim Current Board

Introduction to SCB20

Describes the SCB20 as a unified Shim Current Source, its function and compatibility.

SCB20 Technical Data

Lists technical specifications for the SCB20, including current, voltage, and response times.

SCB20 Configurations

Explains the two standard configurations for BOSS1 and other shim systems.

System Architecture and Overview of SCB20

Presents a block diagram of the SCB20's architecture and its control by the ELCB.

SCB20 Protection Features

Details protection against short circuits and over-temperature for the SCB20.

Measurements Provided for SCB20 Diagnostics

Lists diagnostic checks for status/errors, output measurements, and temperature.

SCB20 Front Panel Connectors and LEDs

Describes the connectors and LEDs on the SCB20 front panel.

Service Software for SCB20

Explains how to access service functions for SCB20 via the Shim Service Web.

Setup and Configuration of Shim Functions

Details how to set up shim functions, load BOSS files, and configure parameters.

View and Manual Modification of Shims

Describes how to view and manually modify shim settings.

SCB20 Part Numbers

Lists the Bruker part number for the SCB20 SHIM CURRENT BOARD.

SCB20 Diagnostic and Troubleshooting

Guides on diagnosing SCB20 issues using the Service Web page.

Troubleshooting Shim System and BOSS File Issues

Provides a flowchart for troubleshooting shim system and BOSS file problems.

GAB/2 Gradient Amplifier

Introduction to AV4 GAB/2

Describes the AV4 GAB/2 as a successor to BSMS/2 GAB/2, supporting gradient functions.

GAB/2 Technical Data

Lists electrical characteristics and notes for the GAB/2 gradient amplifier.

GAB/2 Configurations

Explains configurations for single and 3-channel AV4 GAB/2 installations.

Preemphasis Settings for GAB/2

Details how to define preemphasis parameters using TopSpin or the Service Web.

System Architecture and Overview of GAB/2

Presents a block diagram of the AV4 GAB/2 gradient amplifier board.

Protection Features of GAB/2

Describes protection against short circuits, cables, and over-temperature/current.

Status LEDs on the GAB/2 Front Panel

Explains the meaning of the LEDs (POWER, READY, ERROR, PULSE) on the GAB/2.

Measurements Provided for GAB/2 Diagnostics

Details monitor output and LVDS link data logging for diagnostic purposes.

GAB/2 Control State Machine

Illustrates the state machine steps for GAB/2 to reach operational status.

GAB/2 Web Interface and Service Functions

Explains access to GAB/2 service functions via the ELCB Service Web.

Offset Re-Calibration in the Field

Provides instructions for re-calibrating the GAB/2 offset if dynamic compensation fails.

Preemphasis Setting via Service Web

Details how to set preemphasis parameters using the Service Web interface.

GAB/2 Part Numbers

Lists the Bruker part number for the AV4 GAB/2 GRADIENT AMPL BD.

Troubleshooting Procedure for GAB/2 Gradient Problems

Offers a flowchart for diagnosing and resolving gradient problems with the GAB/2.

L-TRX/L-19F Lock Transceiver

Introduction to L-TRX and L-19F

Describes the integrated L-TRX and optional L-19F units and their functions.

System Architecture and Overview of L-TRX/L-19F

Presents block diagrams illustrating the architecture of the L-TRX and L-19F units.

Function Description of L-TRX/L-19F

Details signal processing, power amplifier control, gradient shimming, and communication interfaces.

Protection Features for L-TRX/L-19F

Explains supervision of power supply, reference clock, temperature, and current.

Internal Diagnostics for L-TRX

Describes L-TRX's extensive internal diagnostic functions accessible via Service Web.

Technical Data for BSMS/2 Lock Transceiver (L-TRX)

Provides detailed technical specifications for the L-TRX transmitter and receiver.

Technical Data for BSMS/2 19F Lock Transceiver

Lists technical specifications for the 19F Lock Transceiver, referencing L-TRX data.

2H Lock with L-TRX Internal Power Amplifier

Details the minimal system for 2H lock using the internal power amplifier.

2H Lock with Additional External Power Amplifier

Explains the use of a more powerful external amplifier for 2H operations.

19F Lock with External 2H Power Amplifier

Illustrates the system setup for 19F Lock with an external 2H power amplifier.

AVANCE NEO OneBay/TwoBay Configurations

Notes compatibility of AV4 BSMS chassis with current generation lock systems.

Front Panel Connectors and LEDs for L-TRX

Describes the connectors and status LEDs on the BSMS/2 Lock Transmitter 300.

Front Panel Connectors and LEDs for L-19F

Details the connectors and status LEDs on the BSMS/2 19F Lock Transceiver.

L-TRX Specific Error Messages

Lists L-TRX error messages and corresponding measures or descriptions.

Part Numbers for L-TRX and L-19F

Provides lists of Bruker part numbers for L-TRX and L-19F lock transceivers.

BSVT Introduction and Configurations

Introduction to the BSVT System

Introduces the BSVT system as a replacement for former VT systems, highlighting modularity.

BSVT Hardware Components

Lists the required hardware components for the BSVT system, including SPB and VPSB boards.

BSVT Software and Features

Details the BSVT software features, control architecture, and supported accessories.

BSVT General Specifications

Outlines general specifications of the BSVT, including architecture and channel capabilities.

BSVT VT Control Electronics Specifications

Provides specifications for temperature setting resolution and applicable temperature range.

Full Electronic VT Gas Control

Details the VT gas control system, flow rates, precision, and pressure monitoring.

Probe Interfaces No Longer Supported

Lists discontinued probe interfaces and their corresponding part numbers.

Basic BSVT Configuration Requirements

Specifies minimal requirements for BSVT configurations, including chassis and ELCB.

Required Boards for BSVT Configuration with VT System Option

Lists required boards based on magnet system type and VT system option.

Support for Nitrogen Level Sensors

Explains support for digital and analog nitrogen level sensors with different BSMS units.

Required Cables for VT Options

Illustrates the necessary cables for connecting VT options to the BSMS configuration.

BSVT Probe Adaptation

Describes how probes and VT accessories are adapted using smart VT interfaces.

HR RT Probes (Thermocouple Type T) Adaptation

Details the adaptation of HR RT probes using a VTA TC-2T.

HR RT Probes (BTO2000) Adaptation

Describes the adaptation of HR RT probes using a VTA BTO.

CryoProbes Adaptation

Explains the connection and safety precautions for CryoProbes with BSVT.

Solids Probes (2 Thermocouple Type T) Adaptation

Details the adaptation of Solids Probes DVT with two thermocouple type T.

BSVT and HT Solids Probe (High Temperature) Adaptation

Explains the direct connection of HT Solids Probe heaters to VPSB-DC-E.

BSVT and HT Accessory (High Temperature) Adaptation

Describes the BVTE3900 cooling system for high temperature NMR.

BSVT and VT Gas Cooling Accessory Adaptation

Details the connection of BCU-I / BCU-II and BCU05 / BCU-X cooling units.

BSVT and Flow Probe Adaptation (FLOW-NMR)

Presents configurations for Flow-NMR probes with HT Heated Probe Capillary.

BSVT Concept and Control Logic

Plug and Play Concept of BSVT

Explains the BSVT's design for seamless integration of probes, chillers, and adapters.

Control Logic of the BSVT System

Describes the common control logic and states (On/Off, Self Test, Errors) of the BSVT.

BSVT State: Off / Standby

Details the state when the BSVT is switched off, including heater, chiller, and gas flow status.

BSVT State: On / Operating

Explains the active state of the BSVT when switched on, including active components.

BSVT State: Sensor Error Handling

Describes how the BSVT enters the sensor error state if a sensor connection is lost.

BSVT State: Gas Flow Error Handling

Explains the causes and states related to gas flow errors in the BSVT.

BSVT State: Self-Test Procedure

Recommends running a self-test if the BSVT or connected devices behave unexpectedly.

Specific Configurations for BSVT

Introduces specific configurations, including BSVT with CryoProbes.

MAS Probes with Tempered Bearing Gas (VTN / WVT)

Details configurations for MAS probes using external VT gas supply.

Frequently Asked Questions about BSVT

Answers common questions regarding BSVT channels, probe connections, and device operation.

SPB (Sensor & Pneumatic Board) Information

Introduction to the SPB

Describes the SPB as a replacement for older boards, supporting standard and wide bore systems.

SPB Configurations (SPB vs. SPB-E)

Details the two SPB variants and their suitability for different system types and features.

Technical Data Overview: SPB vs. SPB-E

Compares SPB and SPB-E interfaces and software-regulated gas flows.

System Architecture and Overview of SPB

Presents block diagrams illustrating the architecture of the SPB.

System Architecture and Overview of SPB-E

Presents block diagrams illustrating the architecture of the SPB-E.

SPB Protection Features

Details protection against short circuits for all external interfaces.

Measurements Provided for SPB Diagnostics

Lists diagnostic checks for status/errors, gas flow, and temperature.

SPB Calibration Procedures

States that there are no calibration settings stored on the SPB; ELCB handles setup.

SPB Front Panel Connectors and LEDs

Describes the connectors and status LEDs on the SPB front panel.

SPB Connectors Description

Details the function and notes for each connector on the SPB.

Function Description: Liquid Helium Level Measurement

Explains how the helium level sensor is monitored using the SPB.

Function Description: Analog Liquid Nitrogen Level Measurement

Describes the analog nitrogen level measurement process using the SPB-E.

Function Description: Sample Down and Sample Up Detection

Details the interface for sample detection using the BST.

Function Description: Version of the Shim Upper Part

Explains how the shim upper part version can be read by the SPB.

Function Description: Sample Changer Interface

Describes the interface for the sample changer's pneumatic controller.

Function Description: Control Output for BSMS/2 Variable Power Supply Board

Details the SPB interfaces for connecting VPSB boards and control signals.

Service Functions for SPB

Explains how to access service functions for the SPB via the ELCB Service Web.

SPB Service Web Page

Provides information about the SPB board and functions controlled by the ELCB.

SPB Diagnostic and Troubleshooting

Guides on diagnosing and troubleshooting issues with the SPB, including LEDs and gas flow.

VPSB-DC and VPSB-DC-E Variable Power Supply Boards

Introduction to VPSB-DC and VPSB-DC-E

Introduces the VPSB-DC as a successor to VPSB, with DC power and backplane connectivity.

Configurations of VPSB-DC and VPSB-DC-E

Describes the two board variants and their capabilities for temperature channels.

Technical Data, Environment, and Norms for VPSB-DC/DC-E

Lists technical data including ambient operating temperature, safety, and protection degree.

Electrical Specification for VPSB-DC

Provides electrical specifications for the VPSB-DC, including output power and voltage.

Electrical Specification for VPSB-DC-E

Details electrical specifications for the VPSB-DC-E, including booster functionality.

System Architecture and Overview of VPSB-DC

Presents a block diagram of the VPSB-DC's architecture.

System Architecture and Overview of VPSB-DC-E

Presents a block diagram of the VPSB-DC-E's architecture.

Control FPGA Functionality

Explains how the Control FPGA receives commands and controls power stages.

Power Supply and Output Power Control

Details output power limits, budgeting, and control for VPSB-DC/DC-E channels.

Protection Features of VPSB-DC/DC-E

Describes protection against short circuits and over-heating of power stages.

Measurements Provided for VPSB-DC/DC-E Diagnostics

Lists diagnostic checks for power supply, connections, and over-current conditions.

VPSB-DC/DC-E Calibration

States that no calibration settings are stored on the VPSB-DC(-E).

Front Panel Connectors and LEDs for VPSB-DC/DC-E

Describes the connectors and status LEDs on the VPSB-DC/DC-E front panel.

Service Functions for VPSB-DC/DC-E

Explains access to service functions for the VPSB-DC/DC-E via the Service Web.

VPSB-DC/DC-E Diagnostic and Troubleshooting

Guides on diagnosing VPSB issues, checking LEDs, and servicing fuses.

VTA (Variable Temperature Adapter) Information

Introduction to VT Adapters (VTA)

Explains the role of VTAs in adapting sensors and heaters for the BSMS VT system.

Configurations of VT Adapters

Lists the various applications and VTA variants presented in subsequent chapters.

List of Available VTAs

Provides a list of available VTAs, their markings, and typical usage.

Technical Data for VTA TC-T

Lists technical specifications for the VTA TC-T, including temperature measurement.

Technical Data for VTA BTO

Lists technical specifications for the VTA BTO, including BTO2000 supply.

Technical Data for VTA CRP

Lists technical specifications for the VTA CRP, including PT100 accuracy.

Technical Data for VTA TC-2T

Lists technical specifications for the VTA TC-2T, including thermocouple type T.

Technical Data for VTA FLOW-NMR

Lists technical specifications for the VTA FLOW-NMR, including thermocouple type T.

Technical Data for VTA TC-2E

Lists technical specifications for the VTA TC-2E, including thermocouple type E.

Technical Data for VTA LN2

Lists technical specifications for the VTA LN2, including level measurement.

Technical Data for VTA BCU

Lists technical specifications for the VTA BCU, including control signal output.

Technical Data for VTA MAG-RS

Lists technical specifications for the VTA MAG-RS, including temperature measurement.

VTA Cable Connectors

Illustrates and labels the various VTA cable connectors.

System Architecture and Overview of VTA

Presents a block diagram of the VT Adapter's system architecture.

Thermocouple Adaptation (VTA TC-2T, TC-2E)

Details thermocouple signal adaptation, cold-junction compensation, and heater filtering.

BTO2000 Adaptation (VTA BTO)

Explains the adaptation of BTO2000 signals and power supply via VTA BTO.

CryoProbe Adaptation (VTA CRP)

Describes CryoProbe adaptation, safety sensor wiring, and connection precautions.

BCU-05 and BCU-X Adaptation (VTA BCU)

Details integrating legacy BCU chillers using the VTA BCU adapter.

Adaptation of Low Temperature Options (VTA LN2)

Explains adaptation of LN2 Heat Exchanger and Evaporator for temperature control.

Protection Features of VTA

Details protection against short circuits and monitoring of temperature and communication.

Measurements Provided for VTA Diagnostics

Explains VTA's detection of shorted/disconnected lines and data reporting.

VTA Calibration

States that no calibration settings are stored on the VTA.

VTA Connectors and LEDs

Shows typical VT adapters and explains the function of LEDs indicating adapter status.

Service Functions for VTA

Explains access to VTA service functions via the VTA Service Web.

VTA Diagnostic and Troubleshooting

Guides on diagnosing VTA issues using the device's 5 LEDs and ELCB data.

Nitrogen Level Sensor

Introduction to the Digital Nitrogen Level Sensor

Introduces the new digital sensor compatible with AVANCE and BSMS BSVT electronics.

Configuration of the Nitrogen Level Sensor

Explains sensor length necessity and system-level configurations (digital/analog mode).

Digital Configuration for AV4 BSMS with BSVT

Illustrates the digital configuration setup for the nitrogen level sensor.

Protection Features of the Nitrogen Level Sensor

Details protection against short circuits and monitoring of power supply and ADC state.

Measurements Provided for Nitrogen Level Sensor Diagnostics

Lists diagnostic checks for communication, ADCs, and sensor interface connectors.

Nitrogen Level Sensor Calibration

States that the sensor is factory calibrated and requires no field calibration.

Nitrogen Level Sensor Connectors and LEDs

Describes the connectors and LEDs on the nitrogen level sensor, indicating level and errors.

Service Functions for Nitrogen Level Sensor

Explains that the sensor is factory calibrated and service is via ELCB Service Web.

Nitrogen Level Sensor Diagnostic and Troubleshooting

Guides on diagnosing issues using the sensor's 5 LEDs and ELCB data.

Radiation Shield Temperature Monitoring (MAG-RS)

Introduction to Radiation Shield Temperature Monitoring

Explains the need for monitoring refrigerated radiation shield temperature in magnet systems.

Configurations and Installation of MAG-RS

Lists the available unit and explains its connection to the BSMS/2 console.

Protection Features for MAG-RS

Details protection against short circuits for external interfaces.

Measurements Provided for MAG-RS Diagnostics

Lists diagnostic checks for communication, ADCs, and power voltages.

MAG-RS Calibration

States that the unit is factory calibrated and requires no field calibration.

MAG-RS Connectors and LEDs

Refers to general VTA connector and LED information.

Service for MAG-RS Unit

States that the unit is factory calibrated and service is via ELCB Service Web.

MAG-RS Diagnostics and Troubleshooting

Guides on diagnosing issues using the device's 5 LEDs and ELCB data.

Installation and Initial Commissioning

Installation Requirements and Guidelines

Specifies that installation must be done by authorized personnel in a controlled environment.

Firmware Check and Download Procedure

Outlines the principle of firmware upgrading, including checking and downloading files.

Firmware Files and Hardware Codes

Explains the role of BsmsCheckDownload.txt, Swiss FTP server, and hardware codes.

Setup of BSMS Firmware

Shows the BSMS Service Web setup screen displaying firmware and hardware configuration.

Operation Guidelines for BSMS System

General Operating Guidelines

Provides general rules for operating the BSMS system, including startup and user interface.

Operator Protection and Safety

Warns about potential dangers from electrical circuitry and improper operation.

Temperature Changes and Device Operation

Advises waiting for the device to reach room temperature after significant ambient changes.

Maintenance Procedures for BSMS

Maintenance Schedule for Laboratory Personnel

Outlines a schedule for daily cleaning and as-needed tasks like firmware updates.

Cleaning Guidelines

Provides instructions for cleaning the device and cabinet units using water or neutral fluids.

Other Cleaning Operations

Advises contacting Bruker Service for special cleaning operations or contamination issues.

Preventative Maintenance Recommendations

Suggests preventative maintenance, especially for the Fan Tray due to fan life expectancy.

Bruker Service Maintenance Options

Informs users about available maintenance options and packages from Bruker.

Software Information and Logging

Explains how device logs are used by customer service for diagnosis.

Firmware Update Procedure

Recommends regularly checking for and performing firmware upgrades.

Troubleshooting BSMS System

General Troubleshooting Steps

Guides on checking power voltages, firmware status, and collecting information for service.

Diagnostic and Troubleshooting Checks

Lists points to check for BSMS problems, including LEDs, firmware, and logging.

Savelogs Command for Information Collection

Explains the Savelogs command for simplifying information collection for customer service.

Replacement of Parts

Personnel Authorized for Part Replacement

States that only Bruker trained or qualified personnel can install or replace parts.

ESD Precautions During Servicing

Advises using ESD discharge bracelets and grounding before touching PCBs.

Loss of Guarantee due to Non-Approved Parts

Warns that using non-approved spare parts invalidates the manufacturer's guarantee.

Procedure for Returning the Unit for Repair

Details the steps for returning a unit for repair, including contacting Bruker and RMA process.

Dismantling and Disposal Procedures

Dismantling Guidelines

Provides steps for safe dismantling, including shutdown, power disconnection, and cleaning.

Disposal Information for Europe (EU)

Offers environmental information for laboratory and industrial customers in the EU regarding WEEE.

Disposal Information for USA and Other Countries

Provides guidance on disposal and recycling regulations and contacts for US and other countries.

Technical Data of the BSMS System

General Information on BSMS System

Provides general data including weight, dimensions, and sound pressure level.

Connection Values for AV4 PSM-48V

Details electrical connection values for the AV4 PSM-48V power supply module.

Connection Values for AV4 PSB-12V

Details electrical connection values for the AV4 PSB-12V power supply module.

Environmental Conditions for Operation

Specifies operating environment conditions like temperature, humidity, and altitude.

Rating Plate Information

Describes the information found on the rating plate of the power supply modules.

Contact Information for Bruker

Bruker Corporation Manufacturer Details

Provides contact address, website, and WEEE information for Bruker BioSpin GmbH.

NMR Hotlines and Service Centers

Offers contact information for Bruker BioSpin NMR service centers and dedicated hotlines.