Bruker NMR BSMS System - Technical Data for VTA TC-2 T; Technical Data for VTA FLOW-NMR

Bruker NMR BSMS System
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VTA
184 / 234 H172203_1_001
VTA TC-2T
Parameter Min Type Max Uni
t
Temperature measurement Range -270 400 °C
Thermocouple type T Accuracy, without
sensor
+/- 0.5 °C
Number of channels 2
Connector type
a
A
Heater Connector type H
Maximum voltage 50 V
Maximum current 7 A
Safety temperature measurement Range -200 +850 °C
Thermocouple type K Resolution 0.1
Measurement update rate 1 s
-1
a
Connector types: See Figure 16.1 [}187]
Table16.5: VTA TC-2T
VTA FLOW-NMR
Parameter Min Type Max Uni
t
Temperature measurement Range -200 350 °C
Thermocouple type T Accuracy (without
sensor)
+/- 0.5 °C
Type 1
Number of channels 1
Connector type
a
A
Heater Connector type J
Maximum voltage 50 V
Maximum current 7 A
Safety temperature measurement Range -200 +850 °C
PT100 Resolution 0.1
Measurement update rate 1 s
-1
a
Connector types: See Figure 16.1 [}187]
Table16.6: VTA FLOW-NMR