Radiation Shield Temperature Monitoring (MAG-RS)
H172203_1_001 203 / 234
18.2.3 Measurements Provided for Diagnostics
The on-board diagnostics supervise essential board functions like power supply and ADC
state. In case of a failure the board will reset and reboot.
The software running on the ELCB may notify the user about abnormal events.
Status/Errors
The MAG-RS unit can perform the following checks:
• Communication and functionality of the ADC’s.
• Power voltages.
• Shorted or disconnected lines at sensor interface connectors.
18.2.4 Calibration
Factory calibration is stored on the board (no field calibration necessary).
18.2.5 Connectors and LED‘s
See Connectors and LED‘s [}190].
18.2.6 Service
This unit does not need any calibration in the field, the product is factory calibrated.
18.2.6.1 Service Web
For a pulse tube replacement or service, a heater must be activated.
There is an ELCB Service Web page available for service. Please contact your local Bruker
office.
18.2.6.2 Diagnostics and Troubleshooting
The device state is displayed using the 5 LED. See LED‘s CH1/2/3/4 and VT POWER [}190]
for a detailed description.
In addition, on-board diagnosis data or failure events are sent to the ELCB immediately and
displayed within TopSpin GUI or logfile.
18.2.7 System Requirements
See Introduction [}201].
18.2.8 Ordering Information
Please contact your local Bruker office.