VTA
182 / 234 H172203_1_001
Bruker
Part
Number
Name Marking Typical usage
[Power or
Auxiliary]
a
Typical usage
Connected devices
Z119237 BSMS/2 VTA TC-2T TC-2T POWER RT, Solids and Flow
Probes
VTN/WVT/DVT
Z116924 BSMS/2 VTA BTO BTO POWER RT probes with BTO2000
Z116923 BSMS/2 VTA CRP CRP POWER CryoProbe
Z120851 BSMS/2 VTA FLOW-NMR FLOW-NMR POWER Flow NMR Capillary
Heater
Z120728 BSMS/2 VTA TC-2E TC-2E POWER Probes for solids, high
temperature
Z119238 BSMS/2 VTA LN2 LN2 POWER N2 Evaporator
N2 heat exchanger
Z116925 BSMS/2 VTA BCU BCU AUX BCU-05, BCU-X
a
POWER means that this application needs heater power and therefore must be connected through
a cable that contains heater power wires to an output of the VARIABLE POWER SUPPLY BOARD
(VPSB-DC(-E)).
Table16.1: List of Available VTAs
16.3 Technical Data
VTA TC-T
Parameter Min Type Max Unit
Temperature measurement Range -270 400 °C
Thermocouple type T Accuracy +/- 0.5 °C
Number of channels 1
Connector type
a
A
Heater Connector type H
Maximum voltage 50 V
Maximum current 7 A
Safety temperature measurement Range -200 +850 °C
Thermocouple type K Resolution 0.1
Measurement update rate 1 s
-1
a
Connector types: See Figure 16.1 [}187]
Table16.2: VTA TC-T