Technical Data
220 / 234 H172203_1_001
25.3 Environmental Conditions
Data Value Unit
Temperature range +5 .. +40 °C
Relative humidity at 31 °C, maximum
a
max. 80 %
Working altitude above sea level max. 2000 m
Supply voltage variation max.-10 ..
+10
%
Overvoltage category II
Temporary overvoltage 1440 V
Pollution degree 2
a
Decreasing linear till relative humidity < 50% at 40 °C.
Table25.4: Operating Environment
For the appropriate temperature see also the Bruker site planning guides.
25.4 Rating Plate
The rating plate is located near the power input of each power supply module and includes
the following information:
Manufacturer
Type
Voltage
Frequency
Apparent power consumption or input current, maximum
PN: Part Number
SN: Serial Number
ECL: Engineering Change Level
Figure25.1: Technical Data: Rating Plate Example