L-TRX/L-19F
H172203_1_001 109 / 234
LOCK IN (J1, SMA):
Receiver input from HPPR/2 or AQS 1H2H preamplifier (LOCK-OUT).
• Maximum input power with no damage: +0 dBm.
2H REC OUT (J2, SMA):
Receiver output to L-TRX 2H-REC IN J1.
REF IN (J3, SMA):
Reference clock input from AV4 REFERENCE board.
• Nominal input power @ 160 MHz: -0.5 ±0.5 dBm.
• Nominal input power @ 320 MHz: -3.5 ±0.5 dBm.
REF OUT (J4, SMA):
Reference clock output to L-TRX REF IN J3.
19F-TR OUT (J7, SMA):
19F RF-output to HPPR/2 19F preamplifier (19F TRANSM).
• 19F Transmitter output specification see Technical Data BSMS/2 19F Lock Transceiver
[}100].
2H-TR-IN (J8, SMA):
Deuterium lock signal from L-TRX 2H-TR OUT J9.
• Maximum input power with no damage: +37 dBm (2H operation).
• Maximum input power with no damage: +10 dBm (19F operation).
2H-TR OUT (J9, N):
2H RF-output to any high power amplifier (FX-IN).
• Transmitter output specification see Technical Data BSMS/2 19F Lock Transceiver
[}100].