GAB/2
94 / 234 H172203_1_001
13. Check the LVDS cables
Verify that only the certified LVDS cables are used, if you are not sure about the required
type, contact your Service support. You might try with a cable from a SGU/2 channel.
15. If the GAB/2 is in error state
Normally, the cause for a GAB/2 error is reported to TopSpin by an error message. One
possible cause could be a missing power supply voltage. It is helpful to check also the
logging provided by the BSMS Service Web (page Service | Display logged messages).
There might be a hint why the GAB/2 is in error state.
Long term BSMS logging can be enabled on the bsmsdisp provided by TopSpin. The enable
check box is on the service page (password required), and the logging can be viewed on the
same page.