CAPINTEC, INC. CAPRAC
®
-t
10-10 WIPE MEASUREMENT PROCEDURES April 2021
Measure
Place the sample for the requested Wipe Location in the Well Counter and then touch
the MEASURE button to begin counting.
The system will begin measuring the Wipe sample. As counting progresses, the live
spectra are shown. The Real time, Live time and the counting rate (total cpm) will be
displayed and updated. Also, the MEASURE button will be replaced with a STOP
button. At any time during the measurement, touch the STOP button to stop the
measurement. To re-start the process, touch the MEASURE button.
When the measurement is finished, the total and ROI counting rates are displayed
and the message “FINISHED” will appear on the screen below the MEASURE button.
Also, a SKIP WIPE and a REPORT button will appear as shown in Figure 10-8 Wipe
Measurement Screen.
Figure 10-8 Wipe Measurement Screen

Table of Contents