RECORDINGS
VF8 DELTA/DELTA XL/KAPPA 7-5
2 ECG Filter M
D
E
Monitor
Off
ESU
3 Pacemaker Detection C
c
On - Artifact Rejection<Medium>
Off - Artifact Rejection<Medium>
4 QRS/ARR processing 2
1
ECG1 + ECG2
ECG1
5 Patient Category/QRS
Classification
<Space>
1
2
B
n
Adult, Neither lead completed learn
Adult, ECG1 lead completed learn
Adult, ECG2 lead completed learn
Adult, ECG1 & ECG2 leads completed learn
Neonate
6 Leads available for
processing
0
1
2
3
No valid lead to process
ECG1 lead valid to process
ECG2 lead valid to process
ECG1 & ECG2 lead valid to process
7 VT Count 5-F Value = VT Count
(where A-F corresponds to 10-15)
8 VT Rate 0-A Value = (VT Rate - 100)/10
(where A corresponds to 10)
9 SVT Count 3-A Value = SVT Count
(where A corresponds to 10)
10 SVT Rate 0-A Value = (SVT Rate - 100)/10
(where A corresponds to 10)
11 TACH Count 5-F Value = TACH Count
(where A-F corresponds to 10-15)
12 TACH Rate 0-A Value = (TACH Rate - 100)/10
(where A corresponds to 10)
13 BRDY Rate 0-F Value = (BRDY Rate - 30)/5
(where A-F corresponds to 10-15)
14 PAUS Rate 0-5 Value = (PAUS Rate - 1.0)/0.5
(where A-F corresponds to 10-15)
15 HR Source E
P
S
ECG is HR Source
IBP(AP) is HR Source
SPO
2
is HR Source
16 RESP Mode O
M
A
Resp Monitoring Off
Manual
Automatic
Position Description Values Definition

Table of Contents