STATUS M ESSAGES
VF8 DELTA/DELTA XL/KAPPA 8-25
ECG Artifact Patient movement, shivering,
tremors
Bad electrode contact
Excessive signal noise
Interference from auxiliary
equipment
Calm the patient.
Check the electrodes and reapply, if
necessary.
Make sure the patient’s skin is properly
prepared.
Isolate the patient from auxiliary
equipment, if possible.
Collecting ECG
Waveforms
The monitor has begun a 10-
second capture of Rest ECG
waveforms.
Instruct the patient to lie still.
ECG Leads
Invalid
The QRS processing leads are
invalid for > 10 sec.
Bad electrode contact or faulty
lead set
Unplugged lead set
The wrong cable type was
selected on the ECG Lead setup
menu (see page 8-20).
Inspect and replace defective cables and
wires.
Reapply gel on disposable electrode(s).
If a lead or electrode cannot be replaced,
select another ECG lead for processing.
Verify that the number of leads selected in
the ECG Lead Setup menu matches the
applied lead set (see page 8-20).
ECG Report
Server Busy-Try
Later
The Infinity CentralStation is
currently processing a report.
Wait a few minutes, then try again.
ECG
H/W Failure
An ECG hardware failure was
detected by an 1 mV test at
startup.
Contact the hospital’s technical personnel
or DrägerService.
MultiMed Pod
Disconnected
The MultiMed pod is not connected
to the monitor during 3, 5, or 6-lead
monitoring.
Check the cables and connection; replace
cables, if necessary.
If the MultiMed pod is not in use, press the
Alarm Silence fixed key.
Message
Definition and/or
Possible Cause
Suggested Action