STATUS M ESSAGES
VF8 DELTA/DELTA XL/KAPPA 8-25
ECG Artifact • Patient movement, shivering,
tremors
• Bad electrode contact
• Excessive signal noise
• Interference from auxiliary
equipment
• Calm the patient.
• Check the electrodes and reapply, if
necessary.
• Make sure the patient’s skin is properly
prepared.
• Isolate the patient from auxiliary
equipment, if possible.
Collecting ECG
Waveforms
The monitor has begun a 10-
second capture of Rest ECG
waveforms.
Instruct the patient to lie still.
ECG Leads
Invalid
• The QRS processing leads are
invalid for > 10 sec.
• Bad electrode contact or faulty
lead set
• Unplugged lead set
• The wrong cable type was
selected on the ECG Lead setup
menu (see page 8-20).
• Inspect and replace defective cables and
wires.
• Reapply gel on disposable electrode(s).
• If a lead or electrode cannot be replaced,
select another ECG lead for processing.
• Verify that the number of leads selected in
the ECG Lead Setup menu matches the
applied lead set (see page 8-20).
ECG Report
Server Busy-Try
Later
The Infinity CentralStation is
currently processing a report.
Wait a few minutes, then try again.
ECG
H/W Failure
An ECG hardware failure was
detected by an 1 mV test at
startup.
Contact the hospital’s technical personnel
or DrägerService.
MultiMed Pod
Disconnected
The MultiMed pod is not connected
to the monitor during 3, 5, or 6-lead
monitoring.
• Check the cables and connection; replace
cables, if necessary.
• If the MultiMed pod is not in use, press the
Alarm Silence fixed key.
Message
Definition and/or
Possible Cause
Suggested Action