RESP SETUP MENU
VF8 DELTA/DELTA XL/KAPPA 12-7
RESP marker Superimposes a vertical line on the
RESP waveform when the monitor
detects a breath. If the monitor
displays the RESP marker because
of artifact or other interference, set
the breath detection threshold to
only count valid breaths:
1. Set the mode to Manual.
2. Enable RESP Marker.
3. Click on Size.
4. Set the Size value at the lowest
value where the RESP marker
appears.
NOTE:
RESP markers are not transmitted
over the network and do not
appear on remote views or
recordings.
The RESP marker indicates the
time of breath detection, not the
beginning or end of respiration.
•ON
•OFF
Coincidence
Detect
Identifies when the respiration rate
is within 20% of the heart rate,
indicating that the monitor may be
counting heart beats as respiration.
WARNING: Respiration/heart rate
coincidence could mask an
apnea condition.
NOTE: Enable respiration alarms
before setting Coincidence Detect
to ON.
ON; The RESP Coincidence message
appears whenever respiration/heart rate
coincidence is detected (default for
Neonatal patient monitoring).
OFF; the monitor does not detect
respiration/heart rate coincidence (default
for adult and pediatric patient monitoring).
RESP
Monitoring
Enables and disables respiration
monitoring. This can also be done
from the Main Screen and ECG
setup menus (see pages 2-5 and 8-
18).
•ON
•OFF
Apnea Time Sets the time the monitor waits
before reporting a cessation of
breathing as an apnea event and
sounds an alarm.
OFF, 10 s, 15 s, 20 s, 25 s, 30 s
NOTE: This feature is available in Neonatal
and Pediatric patient mode only.
Apnea
Archive
Allows user to store and/or record
automatically an apnea alarm
event. User can later review stored
alarms on Event Recall screen.
OFF, Record, Store (default), Str./Rec.
NOTE: This feature is available in Neonatal
and Pediatric patient mode only.
The Respiration Setup Menu
Menu Item Description Settings