STATUS M ESSAGES
VF8 DELTA/DELTA XL/KAPPA 12-19
RESP Lead Off The cause could be one or more
of the following:
Broken cable
Loose lead wire
Faulty lead wire
Dried out gel on electrodes
•M
ULTIMED pod defective
Check the patient cable and lead wires
carefully.
Replace any cable or lead wire that is
suspect.
Reapply gel or change the electrode.
Check the M
ULTIMED pod and replace it, if
necessary.
RESP
Relearning
The user has turned on
respiration monitoring or has
clicked on Relearn.
None required
RESP Artifact Persistent artifact Check the patient cable and lead wires
carefully.
Check the electrode placement. Change
their position, if necessary.
RESP H/W
Failure
Not applicable Call your hospital’s technical personnel or
DrägerService.