STATUS M ESSAGES
VF8 DELTA/DELTA XL/KAPPA 12-19
RESP Lead Off The cause could be one or more
of the following:
• Broken cable
• Loose lead wire
• Faulty lead wire
• Dried out gel on electrodes
•M
ULTIMED pod defective
• Check the patient cable and lead wires
carefully.
• Replace any cable or lead wire that is
suspect.
• Reapply gel or change the electrode.
• Check the M
ULTIMED pod and replace it, if
necessary.
RESP
Relearning
The user has turned on
respiration monitoring or has
clicked on Relearn.
None required
RESP Artifact Persistent artifact • Check the patient cable and lead wires
carefully.
• Check the electrode placement. Change
their position, if necessary.
RESP H/W
Failure
Not applicable Call your hospital’s technical personnel or
DrägerService.
Message Possible Cause Suggested Action