13-10 DELTA/DELTA XL/KAPPA VF8
13 N
ON-INVASIVE BLOOD PRESSURE
Interval Measurements
You can take a series of single measurements at specific intervals. The interval time is
measured from the start of one measurement to the start of the next. To take a series of
measurements:
To start interval measurements
1. Open the NBP menu (see page 13-8).
2. Click on
Interval Time.
3. Dial to the time interval you want to use.
4. Click on the interval to confirm it.
5. Press the
NBP Start/Stop fixed key
If a series of measurements is already in progress, setting a new interval time resets
the timer.
After the first measurement, the NBP parameter box
displays results. A countdown bar indicates the amount of
time left before the next measurement.
You can take additional single or continuous
measurements without affecting the interval cycle. The
minimum interval is 30 seconds between the end of one
measurement and the start of another to allow for
reperfusion of the limb. To stop an interval measurement
in progress, press the
NBP Start/Stop fixed key. The
monitor stops the current measurement and resumes the
cycle on schedule with the next interval measurement.
NOTE: To take the monitor out of interval mode press the NBP Start/Stop fixed
key twice within one second.
1 Time remaining
2 Time elapsed
3 Interval Time