HARDWARE S ETUP
VF8 DELTA/DELTA XL/KAPPA 14-7
Pressure labels that represent two or four monitor pressure channels appear in LCDs
on the Hemo2 and Hemo4 and PiCCO pods (For more information on the PiCCO pod,
see chapter 26). The HemoMed and the MPod – Quad Hemo, which also support four
monitor IBP parameters, does not have LCDs for pressure labels.
Fixed keys, located along the top of each hemodynamic pod, allow you to execute the
following functions:
— “Smart Zero” zeroes all transducers connected to pod and open to
air.
Wedge — Starts a pulmonary wedge pressure measurement (not available
on the PiCCO pod)
.
C.O. Start — Starts a cardiac output measurement.
NOTE: The monitor automatically assigns P1a-d, P2a-d and P3a-d as temporary
pressure labels when pressure transducers are initially connected. These temporary
pressure labels must be relabeled with an appropriate permanent label corresponding to
the pressure site being monitored.
NOTE: C.O. for the PiCCO pod is derived differently than the Hemo2, Hemo4,
HemoMed pods, or the MPod – Quad Hemo.