IBP SETUP
VF8 DELTA/DELTA XL/KAPPA 14-15
connected to an Aux/Hemo connector of the monitor. The letters [a-d] identify the
transducer location on the pod.
To assign a label to a pressure channel
1. Access an IBP setup menu (see page 14-13).
2. Scroll to
Pressure Labels and click.
3. Scroll to the channel you wish to label and click. The first label in the column
on the right side of the menu is highlighted.
4. Scroll to the desired label and click.
5. Repeat steps 3 and 4 to assign other pressure labels.
1
Transducer location letter
2
HemoMed or Y-cable
connector
3
Hemo2, Hemo4, MPod –
Quad Hemo or PiCCO
Kappa
Delta/Delta XL