TPO2/CO2 SETUP
VF8 DELTA/DELTA XL/KAPPA 18-11
To access the tpO
2
/CO
2
setup menu:
Click on the tpO
2
/CO
2
parameter box (if displayed)
or
1. Press the
Menu fixed key.
2. Click on
Patient Setup.
3. Click on Parameters.
4. Scroll to TP or TP* and click.
Quick Reference Table – tpO
2
/CO
2
Setup
NOTE:
Some tpO
2
/CO
2
setup functions are also available on the password-protected
Unit Manager menu (see page 2-14). This allows the unit manager, head nurse,
or physician to config
ure certain critical tpO
2
/CO
2
functions according to the
hospital’s policies and safety considerations.
The tpO
2
/CO
2
and tpO
2
/CO
2
* setup menus do not time-out.
The tpO
2
/CO
2
setup menu
Menu Item Description Settings/Procedures
Site Timer Limits tpO
2
/CO
2
sensor use to a specified
period of time. When time has expired, an
alarm sounds until the heater is shut down
or until you place the sensor in the
calibration chamber, and a message is
displayed on all network devices. The timer
starts to count down the monitoring session
after sensor calibration. A bar graph in the
tpO
2
/CO
2
parameter box indicates time
remaining.
0.5 - 8.0 hr in half-hour increments
•OFF
NOTES: The following Unit Manager
menu items affect Site Timer setting
in the tpO
2
/CO
2
setup menu. See
page 2-14 for detailed information.
If the Site Timer Control setting
has been set to Manager in the Unit
Manager menu, the Site Timer
setting on the tpO
2
/CO
2
setup
menu is ghosted.
If the Timer OFF Control has been
disabled in the Unit Manager menu,
you cannot disable the site timer on
the tpO
2
/CO
2
setup menu.