TPO2/CO2 SETUP
VF8 DELTA/DELTA XL/KAPPA 18-13
Auto
Pressure
Displays the current atmospheric pressure.
NOTE: If the pod’s internal pressure sensor
is not properly calibrated, you cannot
calibrate the tpO
2
/CO
2
sensor in Auto mode
(Auto Pressure displays ***).
Display-only
This function is ghosted if the Atm.
Pressure Mode setting is set to
Auto.
Manual
Pressure
Allows you to enter barometric pressure
manually
NOTE: If Auto Pressure displays ***, enter
the barometric pressure manually and have
the pressure sensor calibrated by a
qualified DrägerService technician.
1. Select Atm. Pressure Mode on
tpO
2
/CO
2
setup menu.
2. Click on Manual.
3. Select the desired setting.
tpO2 Alarm Accesses tpO
2
/CO
2
values in the Alarm
Limits table (see page 5-6).
Not applicable
NOTE: Some tpO
2
/CO
2
setup functions are also available on the Unit Manager menu, where the
unit manager, head nurse, or physician can configure certain critical tpO
2
/CO
2
functions according
to the hospital’s policies and safety considerations. See page 2-14 for information.
The tpO
2
/CO
2
setup menu
Menu Item Description Settings/Procedures