STATUS M ESSAGES
VF8 DELTA/DELTA XL/KAPPA 18-15
TP Cal Required The sensor must be calibrated before
use and after the following:
Change in sensor temperature setting
Change in site timer setting
Change in manually entered
barometric pressure value
Change in auto heater shutdown
setting
Calibrate the sensor.
TP Cal Accepted The sensor calibration was successful. Apply the sensor and proceed with
monitoring.
TP Cal Failed The sensor calibration was
unsuccessful.
Recalibrate the sensor.
TP Barometric
Cal Failed
The pod’s internal barometric pressure
sensor needs calibration.
Contact DrägerService.
TP Comm
Failure
Loss of communication between
monitor and pod.
Remove the sensor from the patient
immediately.
Check the pod and the cable and
connections. If necessary, replace
the cable or pod.
TP Drift The actual sensor temperature differs
from the target temperature setting by
> 0.3 °C.
Make sure the sensor is securely
placed in the fixation ring and ring is
properly attached to the patient.
Replace the sensor if necessary.
TP Cartridge
(pod)
Incompatible
The tpO
2
/CO
2
pod is incompatible with
the monitor.
Replace the pod.
Contact DrägerService.
TP Pods Only
Supported on
AUX1 or AUX2.
TP pod is connected to the wrong
connector at the monitor.
Connect the TP pod to one of the
bottom two pod connectors of the
monitor (Hemo/Aux. or Aux.).
TP Pod H/W
Failure
The tpO
2
/CO
2
hardware failure. Remove the sensor immediately
from the patient.
Check the sensor; the pod and the
pod connections. Replace the sensor
or the pod, if necessary.
Disconnect then reconnect the pod.
Recalibrate the sensor.
If problem persists, call
DrägerService.
Message Possible Cause Suggested Action