ETCO2 SETUP
VF8 DELTA/DELTA XL/KAPPA 19-13
Respiratory
Sweep Speed
Sets waveform sweep speed on screen display. 6.25 mm/s
12.5 mm/s
25 mm/s
50 mm/s
Averaging Sets interval for CO
2
measurements
NOTE: The monitor reports the maximum value
of etCO
2
during the specified sampling interval.
Breath (end-expiration
point)
•10 s
20 s
Instantaneous (pod only
setting) – Represents the
highest CO
2
value per
one-second interval
Atm. Pressure
Mode
Determines automatic or manual setup for
atmospheric pressure compensation
NOTE: Atmospheric pressure compensation
should be set by qualified personnel only.
Consult with your hospital’s technical personnel/
DrägerService before changing the Atm.
Pressure Mode setting to Manual.
•Auto
•Manual
Atm. Pressure This is the current atmospheric pressure.
This function is ghosted if you have selected
Auto under Atm. Pressure Mode.
540 to 800 mmHg (in
increments of 5)
Measurement
Mode
Configures the monitor for mainstream or
sidestream monitoring
•Main
•Side
Respiration Filter
(module only)
Filters signal artifact caused by cardiogenic
oscillations or Baines rebreathing bumps, which
could cause the monitor to display an
erroneously high respiration rate (RRc)
Normal – Disables the
filter
SpecialEnables this
filtering at the expense of
a certain degree of
monitor responsiveness
Apnea Alarm Displays the Apnea Alarm menu. RRc Apnea Time
Apnea Archive
etCO2 Alarm Accesses the Alarm Limits table, where you can
set upper and lower alarm thresholds.See
chapter 5, Alarms, for information about setting
and displaying alarm limits.
Not applicable
Function Description Settings