19-18 DELTA/DELTA XL/KAPPA VF8
19
ETCO2 (END-TIDAL CO2) MONITORING
etCO
2
Adapter
Cal. Failed
The airway adapter calibration
(zeroing in room air) failed.
Make sure the adapter is properly attached to
the sensor and that its windows are clean. If the
problem persists, try another adapter.
etCO2
Adapter Cal.
Required
The sidestream measurement
mode was initiated, requiring air
way adapter cal. (zeroing in
room air) to calibrate pump.
Calibrate the new adapter.
etCO
2
Tubing
Blocked
Sidestream tubing obstructed, or
the filter is clogged.
Clear blockage in the tubing.
Replace the module/pod.
etCO
2
Tubing
Leak
The sidestream tubing has a
leak.
Change the tubing.
etCO
2
module
Disconnected
The etCO
2
module is not
connected to the monitor.
Check cable and connection. Replace cable if
necessary.
etCO
2
module
Incompatible
Corrupt EEPROM. Wrong
software or hardware version
Try a new module.
Consult the hospital’s technical personnel.
NOTE: If the etCO
2
pod is disconnected, the etCO
2
parameter box disappears.
Message Condition Suggested Action