PAW AND VENT SETUP MENUS
VF8 DELTA/DELTA XL/KAPPA 21-9
Function Description Settings
VENT Setup Menu
Respiratory
Mechanics
Displays a list of continuously updated
ventilator parameters.
See page 21-16 for complete list of
associated parameters.
FV Loop Displays a flow/volume loop Not applicable
Flow Scale Sets the size of the flow waveform. 5, 10, 15, 20, 35, 50, 100, 150, or 200
Lmin. The monitor automatically
adjusts lower scale to negative value
Vol Scale Sets the size of the volume waveform. 5, 10, 25, 50, 75, 100, 250, 500, 750,
1000, and 1500 ml
Monitor automatically adjusts lower
scale to 0
Purge Sensor
Purges the tubing of water
condensation and patient secretions.
NOTE: Though the system periodically
purges the breathing circuit
automatically, it may sometimes be
necessary to purge manually.
Not applicable
NOTES:
• Pressure and flow parameters are
frozen during purging.
• Waveforms are not drawn.
MVe Alarm Opens the Alarm Limits table
beginning at MVe setting.
Not applicable
Gas
Compensation
Informs the monitor of ventilation gas
in current use. See page 19-12 for
more information on gas
compensation.
Air, N2O/O2, >60%, Heliox
NOTE: N
2
O/O
2
is available in OR
mode only.
Agent
Concentration
(available in OR
mode only)
Informs the monitor of the anesthetic
agents in ventilation gas being used.
0 - 20 % (increment 1 %)
Waveform
Display
Sets the type of the displayed
waveform.
•Flow
•Vol
• Parameter 1
• Parameter 2
• Parameter 3
Selects up to 3 parameters for display
in the flow/volume parameter box.
PIF, PEF, TVi, TVe, RRv, MVe, TV lk,
None
Clear
Reference
Clears the reference loop from the
main screen.
Not applicable
Loop Draw
Configures the number of displayed
loops.
1, 4
Vent Mini Trend Selects which parameter is displayed
in the Mini Trend when the split screen
is set to 10 or 60 min trends.
TVi s/m, TVe s/m, TVd aw, MVe s/m,
MVe, RRs/m, RRv, Cdyn, C20/Cdyn,
Raw e, PEF, TValv, MValv s/m, VCO2