21-22 DELTA/DELTA XL/KAPPA VF8
21 R
ESPIRATORY MECHANICS
Setting Number of Loops
You can display either a single loop or four consecutive loops.
To set the number of display loops
1. Open the Paw or Vent setup menu (see page 21-7).
2. Scroll to the bottom of the menu and click on the
More arrow to get to the
second page.
3. Click on
Loop Draw.
4. Select
1 or 4 to set the number of display loops.
or
Use the
Loop Draw control key at the bottom of the loop display (“Displaying Loops”
on page 21-21) to set the number of loops.
1
Pressure/Volume
waveform
2
Display or hide grid
3
Pressure scale
4
Set number of loops
displayed
5
Clear reference loop from
screen
6
Save last completed loop
and time stamp
7
Display Flow/Volume loop
8
Set volume

Table of Contents